To adjust a Trimble S series or VX instrument
These steps apply to any Trimble servo or robotic instrument connected to the controller, except a Trimble SX10 or SX12 scanning total station.
If you are connected to a mechanical Trimble total station, you must adjust it using the control panel on the instrument.
- Set up the instrument on a stable surface.
- Make sure that the instrument is accurately leveled and that the compensator is enabled.
-
Tap
and select Instrument / Adjust.
The procedures available in the Adjust screen depend on the connected instrument.
- Perform each adjustment in turn, using the steps below.
Collimation and trunnion axis tilt
On a Trimble total station, you must perform the HA VA collimation and the trunnion axis tilt adjustments together.
NOTE – Final collimation values must be within tolerance of standard values. If not, the instrument needs to be adjusted mechanically. For more information, contact your local Trimble service provider.
-
Position the instrument as follows:
- For HA VA collimation, the instrument must be at least 100 m from the target and the angle to the target must be less than 3° (3.33 gon) from the horizontal plane.
- For trunnion axis tilt, the angle to the target must be at least 30° (33.33 gon) from the VA measured during collimation.
-
Select Collimation & Trunnion axis tilt.
Current adjustment values for the instrument are shown.
-
Tap Next.
-
Aim at the target and take the collimation measurement.
NOTE – Do not use Autolock during collimation or trunnion axis tilt tests.
You must take a minimum of one observation on each face. If you take more than one observation, complete all observations on face 1 first. Between each observation, turn the instrument away and then re‑aim.
- To change face, tap Chg face and observe the same number of observations on face 2 as you observed on face 1.
- When the number of observations is the same for both faces, tap Continue.
-
Aim at the target and take the trunnion axis tilt measurement in the same way you took the collimation measurements.
The current values and the new instrument values are shown.
- Tap Accept.
Autolock collimation
NOTE – Autolock collimation should be performed after the HA VA Collimation adjustment, if available, has been completed.
- Select Autolock collimation.
- Ensure that there are no obstructions between the instrument and the target, which should be at least 100 m apart.
-
Follow the prompts. Press the keys lightly to avoid bumping the instrument.
EDM constant
- Select EDM constant.
- Tap Next.
- Enter an appropriate EDM constant. The available range is ‑9.99 mm to +9.99 mm.
- Tap Store.